Simulator MS400 Multiparameter

Print
Priced From:R 13106,55
Description

PATIENT SIMULATOR MS400
MS400 Multi-Parameter Patient Simulator - ECG, Arrhythmia, IBP, Respiration, Temperature

MS400 Multiparameter Simulator is a portable device, which can provide comprehensive physiological waveform to content general test and train requirement. It can check the monitor rapidly near the bed, also can test the quality warranty of ECG equipment, monitor and remote therapy equipment, At the same time, it can be used as the training of distinguishing from arrhythmia waveforms to the nurses.

Features:
- 3.5-inch color TFT touch-screen, 320*240 pixels, and friendly operation interface.
- Display the output waveforms synchronously, which is convenient for user to compare with the test equipment.
- It integrates the functions of SD socket interface; user can output the custom waveform by this device.
- Built-in templates list to provide the waveform that is expressed by each preset sequence number, which is convenient for searching.
- Two operation interfaces: Chinese and English.
- Operate with keystroke and touch-screen expediently and rapidly.
- Built-in 3.7V/4000mAh (Rechargeable alkaline battery).
- Optical external power: AC 220V, 50Hz or AC 120V, 60Hz

ECG/Arrhythmia:
ECG
- Conventional: All leads have the same voltage with the II-lead as the reference
- Amplitude: 0.5,1.0,1.5 and 2.0mV
- The accuracy of Amplitude: ±2%
- ECG Rate:
 15,20,25,30,40,60,80,100,120,140,160,180,200,220,240,260,280,300,320,340      and 350 BPM
- ECG artifacts: 50 or 60 Hz, muscle, baseline drift and breathing
- The elevation of the ST segment: -0.8 mV to +0.8 mV, the step is 0.1 mV
- The default: 80 BPM,1mV
Arrhythmia
- Atrial: Atrial Fib(coarse), Atrial fib(fine), Atrial Flutter, Sinus Arrhythmia,
 Missed Beat, Atrial
- Tachycardia, Nodal Rhythm, Supraventricular Tachycardia
- Ventricular: Vent Fib (coarse), Vent Fib (fine), Premature Ventricular
  Contraction (PVC) 6/min,
- PVCs 12/min, PVCs 24/min, Frequent Multifocal, Bigeminy, Trigeminy, Paired 
  PVCs, Run 5
- PVCs, Run 11 PVCs, Vent Tachycardia, Asystole
- Premature Beat: Atrial Premature Atrial Contraction (PAC), Nodal PNC, Left 
 Ventricular Focus
- PVC1, Early Left Ventricular Focus PVC1, Right Ventricular Focus PVC1, Early
  Right
- Ventricular Focus PVC1, R on T Left Ventricular Focus, R on T Right 
 Ventricular Focus, Multifocal PVC
- Conduction Defects: First Degree Heart Block, Second Degree Heart Block,
  Third Degree Heart   Block, Right Bundle Branch Block, Left Bundle Branch
  Block  

ECG Performance Testing:
- ECG amplitudeļ¼š0.5,1.0,1.5 and 2.0mV
- The accuracy of Amplitude: ±2%
- Pulse: 30 and 60 BPM, the width of pulse wave is 60 ms
- Square: 0.125 and 2.0 Hz
- Triangle: 2.0 and 2.5 Hz
- Sine: .0.5, 5, 10, 40, 50, 60 and 100 Hz
- ECG Rate:15, 20, 25, 30, 40, 60, 80, 100, 120, 140, 160, 180, 200, 220, 240, 
 260, 280, 300,   320, 340 and 350 BPM
- The accuracy of ECG Rate: ±1%

Respiration:
- Respiration Rate: 0 (Apnea), 15, 20, 30, 40, 60, 80, 100, 120 and 150 R/MIIN
- Respiration Impedance: 0.2, 0.5, 1.0 and 3.0 Ω
- Baseline Impedance:500,1000,1500 and 2000
- Apnea: 12 s, 22 s, 32 s
- ECG simulator with adjustable heart rate/12-lead output
- 33 kinds of Arrhythmia waveforms
- Pacemaker 
- Two IBP channels
- Variety of respiratory impedance,apnea pauses (0 BrPM) of 12, 22 and 32
  seconds.
- Standard ECG calibration
- Export square,triangle and sine with adjustable frequency

IBP:
- The default: 00/00mmHg
- Dynamic IBP
Channel 1: Artery  120/80mmHg
               Radial 120/80mmHg
               Left Ventricle 120/00mmHg
               Right Ventricle 25/00mmHg
Channel 2: Artery  120/80mmHg
               Radial 120/80mmHg
               Left Ventricle 120/00mmHg
               Right Atrium 15/10mmHg
               Right Ventricle 25/00mmHg
               Pulmonary Artery 25/10mmHg
               Left Atrium 14/4mmHg
- Static-Pressure Levels
Channel 1: 0, 80, 160, 240, 320 and 400mmHg
Channel 2: 0, 50, 100, 150, 200 and 250mmHg